BVIT Specs

A full description of how the BVIT detector operates can be found in Siegmund et al. (2008). The current BVIT detector (installed on the SALT in November 2012) has a Super Gen II photocathode whose detection quantum efficiency is ~ 5 times greater than that of the previous S-20 photocathode of the proto-type detector initially mounted on SALT in 2010 (see below).

BVIT Photocathode Quantum Efficiency Comparison

The BVIT instrument consists of a photon counting image tube detector, a shutter/iris,  a pair of filter wheels (containing selectable ND  and B,V, R and H-alpha filters) and command and control electronics (consisting of amplifiers, low and high voltage power supplies, an FPGA data processor and a ground support PC. The detector is cooled using heat sinks and a glycol cooling line to obtain operational temperatures in the  10C to 30C range (dependent on the time of year).

The BVIT instrument on SALT

BVIT Performance Specification

Field of View:                                          1.9 arc min

Spatial Resolution:                                30 microns

Timing Resolution:                               time stamped to 25ns

Filter Selection:                                     B, V, R, H-alpha, ND0 to ND2.0

Maximum Global Count Rate:            1.0 MHz (whole detector)

Maximum Local Count Rate:          100 kHz/ 9 arc sec2 area

BVIT Count Rate Estimator

Enter the apparent magnitude of a source for a particular color and ND filter (don’t forgot to click ‘Click to Edit’!)

Typical BVIT Exposure Time Estimates

Filter          Mag        Source  Cnt rate

V                   13.0        50,300 / sec

V                   14.0        20,100 / sec

V                 16.0         3,200 / sec

V                 18.0         505/ sec

V                 20.0        85 /sec

 

R                  13.0        53,600/ sec

R                 14.0         21,550 / sec

R                 16.0         3,400 / sec

R                 18.0         550/ sec

R                 20.0        90 /sec

Note: These are source count rates detected over a 9 arc sec sq. area on the detector (i.e. typical “seeing area” of a source). The sky background over the same area for a “grey” lunar age is typically 960 cts/ sec (B-band) and 1500 cts/sec (R-band). When integrated over the whole area of the detector the respective total sky background rates are 1,100,000 cts/ sec  and 1,700,000 cts /sec. Thus, the sky background normally limits the allowable total count rate for BVIT and the source limits the  allowable local count rate. Use of the ND filters is normally required for sources brighter than 14th magnitude.

It is IMPORTANT to understand that your choice of filters is dependent both on the maximum source count rate (< 100 kcts/sec) and also on the maximum
global count rate on the detector of < 1 MHz (this can be due to other sources in the field of view and high sky background due to observing near
full moon). Also, the detector should NOT be operated at temperatures > 35C. The glycol cooling should take care of this.